Especificaciones
- Condición
- usado
- Número de stock
- 28542
- Weight
- 11.5 kg (25.4 lbs)
- Line voltage
- 100 – 240 V~, ± 10 %
- Line frequency
- 50 or 60 Hz, ± 5 %
- Power consumption
- 110 VA, 100 W
- Ambient operating temperature
- 4 – 55 °C (39 – 131 °F)
- Ambient non-operating temperature
- -40 – 70 °C (-40 – 158 °F)
- Humidity
- < 95 % r.h. at 40 °C (104 °F)
- Operating altitude
- Up to 3000 m (9842 ft)
- Detection type
- 1024-element photodiode array
- Light source
- Deuterium lamp
- Number of signals
- 8
- Maximum data rate
- 240 Hz (both spectra and signals)
- Drift
- <0.5·10-3 AU/h at 230 nm
- Linear absorbance range
- >2 AU (5 %) at 265 nm, typically 2.5 AU (5 %)
- Wavelength range
- 190 – 640 nm
- Wavelength accuracy
- ±1 nm, self-calibration with deuterium lines
- Wavelength precision
- <±0.1 nm
- Wavelength bunching
- 2 – 400 nm, programmable in 1 nm steps
- Slit width
- 1, 2, 4 , 8 nm
- Diode width
- ~0.5 nm
- Nominal spectral resolution
- < 1 nm
- Max-light cartridge cell (standard)
- 10 mm, V = 1.0 µL
- Max-light cartridge cell (high sensitivity)
- 60 mm, V = 4 µL
- Max-light cartridge ultra low dispersion (uld) cell
- 10 mm, V = 0.6 µL
- Max-light cartridge high dynamic range (hdr) cell
- 3.7 mm, V = 0.9 µL
- Max-light cartridge cell lss
- 10 mm, V = 1.0 µL
- Maximum operating pressure (mop)1
- 70 bar
- Maximum incidental pressure (mip)2
- 150 bar
- Local control
- Agilent Instant Pilot (G4208A) B.02.19 or above
- Eri
- ready, start, stop and shut-down signals
- Categoría
- Cromatografía en USA
- Subcategoría
- Chromatography liquid
- ID de Anuncio
- 97005840
Descripción
Dimensions (height × width × depth): 140 x 396 x 436 mm (5.5 x 15.6 x 17.2 inches)
Short term signal noise (ASTM):
With 10 mm Max-Light cartridge cell: <±3·10-6 AU/cm at 230 nm, slit width 4 nm, TC 2 s
With 60 mm Max-Light cartridge cell: <±0.6·10-6 AU/cm at 230 nm, slit width 4 nm, TC 2 s
Time programmable: Wavelength, polarity, peak width, lamp bandwidth, auto balance, wavelength range, threshold, spectra storage mode
Flow cells:
User-exchangeable, self-aligning cartridge cells with RFID tags.
User-exchangeable, self-aligning cartridge cells with RFID tags.
User-exchangeable, self-aligning cartridge cells with RFID tags.
Spectral tools: Data analysis software for spectra evaluation, including spectral libraries and peak purity functions
Analog output: Recorder/integrator: 100 mV or 1 V, output range 0.001 – 2 AU, one output
Instrument control:
Lab Advisor B.02.06 or above
LC and CE Drivers A.02.11 or above
For details about supported software versions refer to the compatibility