- System requirements (computer)
- Memory: 512 MB or more Hard disk: 1 GB or more of available space Display: 1,024 × 768 dots or more, 65,536 colors or more Interface: USB 2.0 or more to be installed OS: Windows 8.1 (32-bit/64-bit) Windows 10 (32-bit/64-bit) Windows 11 (64-bit) .NET Framework: 4.7.2
- Login function
- The login screen is displayed when " is accessed The analysis related screen (my page) is displayed when you log in using the GENNECT account
- License authentication
- 3-day license, 30-day license, 365-day license The analysis function is only available when the license is activated
- Analysis item
- - DCR - Rratio - Uniformity
- Impedance data file upload
- - Maximum number of files to be uploaded simultaneously: 30 - Total size limit of files to be uploaded: 1 MB - File extension: CSV Files uploaded simultaneously have the same material information and analysis conditions.
- Supported file formats
- CSV file obtained by the measurement software (SA2633)
- Analysis time (representative value)
- - Approx.1 minute/file - Number of files to be simultaneously analyzed: 50
- Language
- Japanese, English (default setting: English)
- Instrument to be controlled
- IM3536 LCR Meter, IM3536-01 LCR Meter
- Measurement data save function
- Outputs the measured impedance data in the file format that can be loaded using the SA2632 Analytical Software. File extension: CSV
- Measurement time
- Approx. 1 minute
- Dimensions
- Approx. 98W × 38H × 24D mm (3.86″W × 1.50″H × 0.94″D) (excluding protruding parts)
- Weight
- Approx. 210 g (7.4 oz.)
- Material
- Container: polypropylene (PP), electrode: brass (nickel plated)
- Capacity
- Approx. 1 mL
- Electrode pin
- Diameter (Area where liquid to be measured comes in contact): 3 mm ±0.1 mm Electrode interval: 6 mm ±0.3 mm
- Operating temperature and humidity range
- 23°C ±5°C (73°F ±9°F), 80% RH or less (non-condensing)
- Storage temperature and humidity range
- −10°C to 50°C (14°F to 122°F), 80% RH or less (non-condensing)
- Operating environment
- Indoors, altitude up to 2000 m (6562 ft.)
- Product warranty duration
- The Electrode Cell is a consumable and is not covered by the warranty
- Measurable frequency
- DC to 10 MHz
- Connectable sample
- SA9001 Electrode Cell
- Residual impedance
- Residual resistance during short circuit 200 mΩ or less (reference for 100 Hz) Inter-electrode stray capacitance 0.2 pF or less (reference for 1 MHz)
- Included accessory
- Shorting plate for compensation
- Frequency sweep range
- Minimum frequency: 0.9 mHz to 110 kHz Maximum frequency: 190 kHz to 310 MHz
- Number of measurement points
- Between 490 points and 1100 points in logarithmic intervals
- Applied signal
- Constant-voltage, ±100 mV
- Measurement parameters
- Frequency, Rs (ReZ), X (ImZ)
- Active materials
- NMC, LCO, NCA, LFP, LMO, MnO2/CFx, Graphite, LTO, Si, SiO, Activated Carbon, None
- Conductive aid
- Acetylene Black, Furnace Black, Carbon Nanotube, Graphite
- Binder
- PVDF, SBR, PTFE, None
- Dispersant
- CMC, MC, PVP, None
- Solvent
- NMP, Water
- Categoría
- Analizadores de impedancia en Japón
- Subcategoría
- Advanced technologies
- ID de Anuncio
- 71287810
Descripción
2022
HIOKI offers advanced quality test solutions for electrode Slurry.
-Three parameters indicate the mixture condition of the slurry based on impedance
-Hardware & Software ideal for slurry impedance measurement
-Data & Analysis tool for various types of slurry
Unique HIOKI Slurry Analysis technology provides quantified mixing quality of electrode slurry to approach your ideal Battery. It enables you to optimize the slurry process and helps detect quality deviations in future battery cell manufacturing upstream processes. Ultimately, this will lead to the stabilization of cell quality.
A proprietary Hioki algorithm analyzes impedance measured values for LiB electrode slurries
Easily measure impedance of slurry
Analysis Results “DCR, Rratio, Uniformity” indicate electron conductivity of Slurry
Cloud-based analysis software ensures you always have access to the latest version
Able to choose license plan, fit the right solution for your needs